1. Comply with JJF-1190-2008 dust particle counter verification specifications:
2. Self-purification time ≤10min
3. Measurement repetition relative deviation ≤±10%
4. Particle size distribution error 0.5um 5um≤±20%
5. Particle concentration indication error 0.5um≤±20%
6. Sampling flow and error: 2.83L/min ≤±5%
7. Measurable particle size range: 0.3, 0.5, 1, 3, 5, 10um, divided into 6 particle size channels,
signal-to-noise ratio of 0.3um ≥ 2.5:1
8. Maximum measurable particle concentration: ≤30K particles/L
9. Current DC7.2V, 1800mAh
Model Y09-3016E
Light source Laser diode
Flow rate 2.83L/MIN
Particle channels 0.3,0.5,1,3,5,10um / 6 channels
Sampling time User-defined (1 minute to 99 minutes)
Sampling delay User-defined (1 minute to 99 minutes)
Sampling frequency 1 to 99 times or continuously
Zero count ≤10min
Alarm level CClass100~Class300,000 (FED209E standard)
lass 5~9 (ISO14464-1 standard)
Display LCD touch screen
Power AC 220V 60HZ 6V/2500mAH
Printer Bluetooth printer
Storage capacity 1000 groups data
Printer External bluetooth printer
Dimension 220*40*120mm
Weight 1KG
UCL judgment criteria ISO14644 FS209E
Using environment SWorking environment: 15~35℃, 10-85%RH
torage environment: -10~50℃, 10-85%RH